Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Rustrak RANGER II RR2-1235: (1 year) CAL VER /5520 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 19 NUMBER OF LINES: 163 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: -10 - 60 degC 1.004 DISP Humidity: 5 - 90%RH 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Connect the signal input cable to the input 1.006 DISP connector on the side of the Ranger II. 1.006 DISP Be sure to secure it tightly. 1.007 DISP Press BOTH buttons to turn the UUT ON and get to the 1.007 DISP top menu level. 1.007 DISP Press key until top line reads "DISPLAY". 1.007 DISP Press to enter the DISPLAY MODE. 1.008 HEAD {CHANNEL 1: 0 - 600V AC} 1.009 DISP If Ch 1 reading is not displayed, use the LEFT function 1.009 DISP button to "scroll" through available channels until 1.009 DISP Ch 1 is displayed. 1.010 DISP Connect the 5520A to the UUT Ch1 (red) as follows: 1.010 DISP [32] 5520A NORMAL HI to CHANNEL 1 HI INPUT 1.010 DISP [32] 5520A NORMAL LO to CHANNEL 1 LO INPUT 1.011 5520 600 100.00V 0.05/ 60H SI 2W 2.001 5520 600 300.00V 0.05/ 60H SI 2W 3.001 5520 600 600.00V 0.05/ 60H SI 2W 4.001 HEAD {CHANNEL 2: 0 - 600V AC} 4.002 DISP If Ch 2 reading is not displayed, use the LEFT function 4.002 DISP button to "scroll" through available channels until 4.002 DISP Ch 2 is displayed. 4.003 DISP Connect the 5520A to the UUT Ch2 (black) as follows: 4.003 DISP [32] 5520A NORMAL HI to CHANNEL 2 HI INPUT 4.003 DISP [32] 5520A NORMAL LO to CHANNEL 2 LO INPUT 4.004 5520 600 100.00V 0.05/ 60H SI 2W 5.001 5520 600 300.00V 0.05/ 60H SI 2W 6.001 5520 600 600.00V 0.05/ 60H SI 2W 7.001 HEAD {CHANNEL 3: 0 - 600V AC} 7.002 DISP If Ch 3 reading is not displayed, use the LEFT function 7.002 DISP button to "scroll" through available channels until 7.002 DISP Ch 3 is displayed. 7.003 DISP Connect the 5520A to the UUT Ch3 (blue) as follows: 7.003 DISP [32] 5520A NORMAL HI to CHANNEL 3 HI INPUT 7.003 DISP [32] 5520A NORMAL LO to CHANNEL 3 LO INPUT 7.004 5520 600 100.00V 0.05/ 60H SI 2W 8.001 5520 600 300.00V 0.05/ 60H SI 2W 9.001 5520 600 600.00V 0.05/ 60H SI 2W 10.001 HEAD CURRENT MEASUREMENT 10.002 DISP To verify the current measurement function of the 10.002 DISP Rustrak Ranger II, the SCALE for CHANNELS 5-7 10.002 DISP must be set to 1000:1 (DEFAULT SETTING). 10.003 HEAD {CHANNEL 5: 0 - 5A AC} 10.004 DISP If Ch 5 reading is not displayed, use the LEFT function 10.004 DISP button to "scroll" through available channels until 10.004 DISP Ch 5 is displayed. 10.005 DISP Connect the 5520A to the UUT Ch5 (red) as follows: 10.005 DISP [32] 5520A AUX HI to CHANNEL 5 HI INPUT 10.005 DISP [32] 5520A AUX LO to CHANNEL 5 LO INPUT 10.006 5520 1.0000A 60H SI S 2W 10.007 ACC 200.00A 0.22U 10.008 MEMI Enter the UUT reading in amps. 10.009 MEME 10.010 MEMC 1000 A 0.05/ 11.001 5520 2.5000A 60H SI S 2W 11.002 ACC 500.00A 0.65U 11.003 MEMI Enter the UUT reading in amps. 11.004 MEME 11.005 MEMC 1000 A 0.05/ 12.001 DISP Connect the 5520A to the UUT Ch5 (red) as follows: 12.001 DISP [32] 5520A 20A to CHANNEL 5 HI INPUT 12.001 DISP [32] 5520A AUX LO to CHANNEL 5 LO INPUT 12.002 5520 5.0000A 60H SI HC S 2W 12.003 ACC 1000.00A 0.9U 12.004 MEMI Enter the UUT reading in amps. 12.005 MEME 12.006 MEMC 1000 A 0.05/ 13.001 HEAD {CHANNEL 6: 0 - 5A AC} 13.002 DISP If Ch 6 reading is not displayed, use the LEFT function 13.002 DISP button to "scroll" through available channels until 13.002 DISP Ch 6 is displayed. 13.003 DISP Connect the 5520A to the UUT Ch6 (black) as follows: 13.003 DISP [32] 5520A AUX HI to CHANNEL 6 HI INPUT 13.003 DISP [32] 5520A AUX LO to CHANNEL 6 LO INPUT 13.004 5520 1.0000A 60H SI S 2W 13.005 ACC 200.00A 0.22U 13.006 MEMI Enter the UUT reading in amps. 13.007 MEME 13.008 MEMC 1000 A 0.05/ 14.001 5520 2.5000A 60H SI S 2W 14.002 ACC 500.00A 0.65U 14.003 MEMI Enter the UUT reading in amps. 14.004 MEME 14.005 MEMC 1000 A 0.05/ 15.001 DISP Connect the 5520A to the UUT Ch6 (black) as follows: 15.001 DISP [32] 5520A 20A to CHANNEL 6 HI INPUT 15.001 DISP [32] 5520A AUX LO to CHANNEL 6 LO INPUT 15.002 5520 5.0000A 60H SI HC S 2W 15.003 ACC 1000.00A 0.9U 15.004 MEMI Enter the UUT reading in amps. 15.005 MEME 15.006 MEMC 1000 A 0.05/ 16.001 HEAD {CHANNEL 7: 0 - 5A AC} 16.002 DISP If Ch 7 reading is not displayed, use the LEFT function 16.002 DISP button to "scroll" through available channels until 16.002 DISP Ch 7 is displayed. 16.003 DISP Connect the 5520A to the UUT Ch7 (blue) as follows: 16.003 DISP [32] 5520A AUX HI to CHANNEL 7 HI INPUT 16.003 DISP [32] 5520A AUX LO to CHANNEL 7 LO INPUT 16.004 5520 1.0000A 60H SI S 2W 16.005 ACC 200.00A 0.22U 16.006 MEMI Enter the UUT reading in amps. 16.007 MEME 16.008 MEMC 1000 A 0.05/ 17.001 5520 2.5000A 60H SI S 2W 17.002 ACC 500.00A 0.65U 17.003 MEMI Enter the UUT reading in amps. 17.004 MEME 17.005 MEMC 1000 A 0.05/ 18.001 DISP Connect the 5520A to the UUT Ch7 (blue) as follows: 18.001 DISP [32] 5520A 20A to CHANNEL 7 HI INPUT 18.001 DISP [32] 5520A AUX LO to CHANNEL 7 LO INPUT 18.002 5520 5.0000A 60H SI HC S 2W 18.003 ACC 1000.00A 0.9U 18.004 MEMI Enter the UUT reading in amps. 18.005 MEME 18.006 MEMC 1000 A 0.05/ 19.001 END